Menu

Micropette Plus, Multichannel, Fully Autoclavable DLAB

• Dispensing head rotates for effortless pipetting convenience
• Individual piston and tip cone assembly allow for easy repair and maintenance
• Compound material-made tip cone secures high sealing performance
• Compatible with most universal tip brands

SKU: n/a.

Clear

Description

8 Channel Adjustable Volume Pipettes

Volume Range Increment Test Volume(μl) Accuracy Error Precision Error
0.5-10μL 0.1μL 10μL 1.50% 0.15ul 1.50% 0.15ul
5μL 2.50% 0.125ul 2.50% 0.125ul
1μL 4.00% 0.04ul 4.00% 0.04ul
5-50μL 0.5μL 50μL 1.00% 0.5ul 0.50% 0.25ul
25μL 1.50% 0.375ul 1.00% 0.25ul
5μL 3.00% 0.15ul 2.00% 0.1ul
50-300μL 5μL 300μL 0.70% 2.1ul 0.25% 0.75ul
150μL 1.00% 1.5ul 0.50% 0.75ul
50μL 1.50% 0.75ul 0.80% 0.4ul


12 Channel Adjustable Volume Pipettes

Volume Range Increment Test Volume(μl) Accuracy Error Precision Error
0.5-10μL 0.1μL 10μL 1.50% 0.15ul 1.50% 0.15ul
5μL 2.50% 0.125ul 2.50% 0.125ul
1μL 4.00% 0.04ul 4.00% 0.04ul
5-50μL 0.5μL 50μL 1.00% 0.5ul 0.50% 0.25ul
25μL 1.50% 0.375ul 1.00% 0.25ul
5μL 3.00% 0.15ul 2.00% 0.1ul
50-300μL 5μL 300μL 0.70% 2.1ul 0.25% 0.75ul
150μL 1.00% 1.5ul 0.50% 0.75ul
50μL 1.50% 0.75ul 0.80% 0.4ul

DLAB specifications are used as guidelines and the user calibration should refer to the industrial standard ISO 8655.
Error limits in accordance with ISO8655-2