Creates a temperature, humidity, light environment for drug stability evaluation. Suitable for accelerated testing, long-term testing, wet test, and strong light irradiation test which are the best choice for pharmaceutical companies to conduct drug stability tests.
Ergonomic design Continuous Operation Quality Assurance Lighting system The source system can be set with different light intensity at each partition level, and the light intensity can be adjusted between 0 and 100%.
Type
Medicine stability test chamber
Medicine stability test chamber
Medicine stability test chamber
Medicine stability test chamber
Comprehensive medicine light stability test chamber (UV)
Comprehensive medicine light stability test chamber (UV)
Comprehensive medicine light stability test chamber (UV)
Comprehensive medicine light stability test chamber (UV)
Model
BHM-60P
BHM-120P
BHM-250P
BHM-500P
BHM-60GPUV
BHM-120GPUV
BHM-250GPUV
BHM-500GPUV
Temperature Range
-10-80°C
-10-80°C
-10-80°C
-10-80°C
No light -10-80°C, with light 10-70°C
No light -10-80°C, with light 10-70°C
No light -10-80°C, with light 10-70°C
No light -10-80°C, with light 10-70°C
Capacity
72L
120L
248L
495L
72L
120L
248L
495L
Temperature
fluctuation /
Uniformity±0.5°C/±2°C
±0.5°C/±2°C
±0.5°C/±2°C
±0.5°C/±2°C
±0.5°C/±2°C
±0.5°C/±2°C
±0.5°C/±2°C
±0.5°C/±2°C
Temperature Accuracy
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
Humidity range
30%~95%RH/±3%RH
30%~95%RH/±3%RH
30%~95%RH/±3%RH
30%~95%RH/±3%RH
30%~95%RH/±3%RH
30%~95%RH/±3%RH
30%~95%RH/±3%RH
30%~95%RH/±3%RH
Sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Temperature: PT100 Humidity: Imported humidity sensor
Light
–
–
–
–
White fluorescence + UV
White fluorescence + UV
White fluorescence + UV
White fluorescence + UV
Spectral Range
–
–
–
–
UV @320nm~400nm Peak Wavelength @365nm
UV @320nm~400nm Peak Wavelength @365nm
UV @320nm~400nm Peak Wavelength @365nm
UV @320nm~400nm Peak Wavelength @365nm
Light Intensity
–
–
–
–
White fluorescence 0-6000 lux / ± 500 lux, UV light 0-5 W / m2
White fluorescence 0-6000 lux / ± 500 lux, UV light 0-5 W / m2
White fluorescence 0-6000 lux / ± 500 lux, UV light 0-5 W / m2
White fluorescence 0-6000 lux / ± 500 lux, UV light 0-5 W / m2
Light Source
–
–
–
–
White fluorescent lamp (ISO10977 compliant), UVA ultraviolet
fluorescent lampWhite fluorescent lamp (ISO10977 compliant), UVA ultraviolet
fluorescent lampWhite fluorescent lamp (ISO10977 compliant), UVA ultraviolet
fluorescent lampWhite fluorescent lamp (ISO10977 compliant), UVA ultraviolet
fluorescent lamp
Standard Light Source
–
–
–
–
White fluorescent lamp (D65, 6500K), UVA @340nm~400nm UV
lampWhite fluorescent lamp (D65, 6500K), UVA @340nm~400nm UV
lampWhite fluorescent lamp (D65, 6500K), UVA @340nm~400nm UV
lampWhite fluorescent lamp (D65, 6500K), UVA @340nm~400nm UV
lamp
Standard Compliants
GMP
GMP
GMP
GMP
GMP, ICH-Q1B Directive
White fluorescence ≥ 1.2x106Lux.hr, near ultraviolet fluorescence
≥ 200w.hr/m2GMP, ICH-Q1B Directive
White fluorescence ≥ 1.2x106Lux.hr, near ultraviolet fluorescence
≥ 200w.hr/m2GMP, ICH-Q1B Directive
White fluorescence ≥ 1.2x106Lux.hr, near ultraviolet fluorescence
≥ 200w.hr/m2GMP, ICH-Q1B Directive
White fluorescence ≥ 1.2x106Lux.hr, near ultraviolet fluorescence
≥ 200w.hr/m2
Compressor
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
Two sets of independent original fully enclosed compressors
No of Partition
2
3
3
4
2
3
3
4
Internal Dimension
WxDxH400x400x450
500x400x600
550x600x750
670x725x1020
400x400x450
500x400x600
550x600x750
670x725x1020
External Dimension
WxDxH570x770x1120
670x790x1270
720x990x1420
850x1095x1935
570x770x1120
670x790x1270
720x990x1420
850x1095x1935
Power
220V 50HZ
2000W220V 50HZ
2250W220V 50HZ
2500W220V 50HZ
3910W220V 50HZ
2000W220V 50HZ
2250W220V 50HZ
2500W220V 50HZ
3910W
• New fluorine-free design: high efficiency, low energy consumption, and energy saving, so that you are always at the forefront of healthy living.
• Microcomputer controller: intelligent touch control, real-time display of parameters, simple and convenient operation, stable, accurate and reliable control.
• The inner body of the box is made of 304 stainless steel, with a four-corner semi-circular transition design, smooth and easy to clean
• Unique air duct circulation: Ensure that the wind speed inside the studio is evenly distributed.
• A test hole with a diameter of 25mm is standard on the left side of the cabinet.
• Two sets of compressors automatically switches to ensure that the drug test does not malfunction after continuous operation for a long time.
• Continuous operation does not require defrosting to avoid fluctuations in temperature and humidity in the tank during use.
• Quality key components such as temperature and humidity controller, compressor and circulating fan are all imported and installed to ensure
stable operation under long duration.
• The illumination system includes visible light tubes and ultraviolet light tubes, which can control the visible light tube and the ultraviolet light tube separately, or can choose single layer or double-layer visible light tube or ultraviolet light tube
• Lighting system meets ICH Q1A and Q1B related standards, visible light (VIS), ultraviolet (UV) and mixed light