Menu

Large Direct-Pure® EDI, Rephile

High-flow water systems designed to produce Type II water from tap water directly.
Resistivity of the product water is above 5 MΩ·cm at 25 ℃ , which meets or exceeds Type II water quality as defined by ASTM, CAP, CLSI and ISO 3696 / BS 3997 and also complies with the Purified Water requirements from the European and U.S. Pharmacopoeia. The system is CE and RoHS certified.

SKU: n/a.

Clear

Description

<strong>Product No</strong><strong>Specification</strong><strong>Download</strong><strong>Application</strong><strong>Features</strong>
Product Number Description
RD0E0300E EDI 30L, 230 V/50 Hz
RD0E0600E EDI 60L, 230 V/50 Hz
RD0E01H0E EDI 125L, 230 V/50 Hz
 Feed Water Requirements  
 Tap water*  Municipal water, TDS < 1000 ppm (2000 μS/cm)
 Operating Temperature  5~45 ℃
 Feed Water Pressure  30 – 90 psi (2 – 6 bar)
 Product Water Flow Rate  Typically 30, 60 or 125 L/h (at 25 ℃)
 Product Water Quality  
 Resistivity  > 5 MΩ·cm @ 25 ℃ (typically > 10 MΩ·cm @ 25 ℃)
 TOC  < 30 ppb
 Particle (> 0.2 μm)  < 1/ml (with a 0.2 μm final filter)
 Microorganisms  < 0.1 cfu/ml (with a 0.2 μm final filter)
 Pyrogens (endotoxins)  < 0.001 Eu/ml (with a terminal ultrafiltration cartridge)
 Dimensions  
 Length x Depth x Height  40.0 cm X 49.0 cm X 110.0 cm

Type II pure water can be used in many areas.
Some of the typical applications are:

• Preparation of chemical and bio-reagents

• Preparation of culture media

• Preparation of solutions for chemical analysis such as HPLC and ICP

• For clinical analyzers

• Medical device and equipment rinsing

• For serum and blood fractionation

• For ophthalmics

• Feed water for ultrapure water systems

The system incorporates the best in class Siemens EDI module and stringently selected RO membrane, to deliver large amount of quality Type II water consistently at a relatively low cost.

• System has automatic water quality control to ensure EDI water meet or exceed Type II water standards.

• Fully automated control system has pre-rinse, RO membrane fouling preventing flush, RO membrane life detection and many other functions.

• The pre-rinsing program starts automatically once a new RO membrane is installed so there is no need for user to set additional cleaning procedures.

• Product water can be dispensed from the water system directly.

• The system is upgradable to produce both Type II and Type I water by installing an ultra purification cartridge inside the system.

• RephiLe can provide full document support to meet user’s GMP, GLP, FDA and other certification requirements.